• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH EXETER V40 STEM 44MM NO 0; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH EXETER V40 STEM 44MM NO 0; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED Back to Search Results
Catalog Number 0580-1-440
Device Problems Adverse Event Without Identified Device or Use Problem (2993); Insufficient Information (3190)
Patient Problems No Information (3190); Insufficient Information (4580)
Event Date 12/10/2020
Event Type  Injury  
Manufacturer Narrative
It was reported that the patient has not yet been revised.No other information is available at this time.Should additional information become available it will be reported in a supplemental report upon completion of the investigation.Not returned to the manufacturer.
 
Event Description
The customer reported that a patient who had been implanted with an exeter mitch combination needs revision.
 
Manufacturer Narrative
An event regarding revision for unspecified reason involving an exeter stem was reported.The event was not confirmed.Method & results: product evaluation and results: the stem was returned for evaluation.The damage on the stem is consistent with implantation and explantation processes.Nothing else remarkable was noticed.Clinician review: no medical records were received for review with a clinical consultant.Product history review: review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.Complaint history review: there have been no other similar events for the lot referenced.Conclusions: it was reported that the patient's hip was revised for unspecified reason.The exact cause of the event could not be determined because insufficient information was provided.Further information such as pathology reports, pre- and post-operative x-rays and the primary operative report as well as patient history and follow-up notes are needed to complete the investigation for determining root cause.No further investigation for this event is possible at this time.If additional information becomes available to indicate further evaluation is warranted, this record will be reopened.Catalog numbers and lot codes of oem devices listed in this report: cat# mac-9988-4854; std mitch trh cp sz 48/54; lot# fm59166018.Cat# mmh-9988-0048; mitch trh md hd sz 48+0; lot# fm80523.
 
Event Description
The customer reported that a patient who had been implanted with an exeter mitch combination needs revision.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EXETER V40 STEM 44MM NO 0
Type of Device
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
MDR Report Key11119166
MDR Text Key225279439
Report Number0002249697-2021-00036
Device Sequence Number1
Product Code JDI
Combination Product (y/n)N
PMA/PMN Number
K121308
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,other
Type of Report Initial,Followup
Report Date 07/21/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date10/31/2011
Device Catalogue Number0580-1-440
Device Lot NumberG1567951
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/29/2021
Initial Date Manufacturer Received 12/10/2020
Initial Date FDA Received01/05/2021
Supplement Dates Manufacturer Received06/24/2021
Supplement Dates FDA Received07/21/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Other; Required Intervention;
Patient Age56 YR
-
-