• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) ASSURITY RF DR; IMPLANTABLE PACEMAKER PULSE GENERATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) ASSURITY RF DR; IMPLANTABLE PACEMAKER PULSE GENERATOR Back to Search Results
Model Number PM2240
Device Problem Pacing Problem (1439)
Patient Problems Ventricular Tachycardia (2132); Dizziness (2194)
Event Date 04/09/2019
Event Type  malfunction  
Manufacturer Narrative
The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Event Description
It was reported that the patient presented in clinic.Upon interrogation, it was noted that there was a non-sustained ventricular tachycardia (nsvt) episode.The nsvt episode coincided with a ventricular paced event that was followed by the autocapture backup pulse.The backup pulse was triggered by fusion caused by premature ventricular contractions.The technician believed that the autocapture caused the nsvt to occur.Also, the patient had dizzy spells that were due to the nsvt.Programming changes were made to turn off autocapture.The patient was in stable condition.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ASSURITY RF DR
Type of Device
IMPLANTABLE PACEMAKER PULSE GENERATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
sonali arangil
15900 valley view court
sylmar, CA 91342
MDR Report Key11122006
MDR Text Key225252354
Report Number2017865-2021-00748
Device Sequence Number1
Product Code NVZ
UDI-Device Identifier05414734507073
UDI-Public05414734507073
Combination Product (y/n)N
Reporter Country CodeAS
PMA/PMN Number
P880086
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 01/05/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/30/2016
Device Model NumberPM2240
Device Catalogue NumberPM2240
Device Lot NumberP000008390
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/17/2020
Initial Date FDA Received01/06/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/15/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-