Model Number N/A |
Device Problem
Difficult to Insert (1316)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 12/17/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).Report source: (b)(6).Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
|
|
Event Description
|
It was reported that during an initial hip arthroplasty, the liner would not seat into the shell.No adverse events have been reported as a result of the malfunction and additional information on the reported event is unavailable.
|
|
Event Description
|
No further event information available at the time of this report.
|
|
Manufacturer Narrative
|
This follow-up report is being submitted to relay additional information. .
|
|
Event Description
|
No further event information available at the time of this report.
|
|
Manufacturer Narrative
|
This follow-up report is being submitted to relay additional information.Reported event was unable to be confirmed.A review of the device history records identified deviations or anomalies during manufacturing, the deviations or anomalies would not have attributed to the event.Root cause could not be determined as the necessary information to adequately investigate the reported event was not provided.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.H3 other text: device not returned for evaluation.
|
|