C.R. BARD, INC. (BASD) -3006260740 POWERPORT ISP M.R.I. IMPLANTABLE PORT, GROSHONG SINGLE-LUMEN, 8F; PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR Back to Search Results
Model Number 9808560
Device Problems Fracture (1260); Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Air Embolism (1697)
Event Date 12/01/2020
Event Type  Injury  
Manufacturer Narrative
As the lot number for the device was not provided, a review of the device history records could not be performed.The return of the sample is pending.The investigation of the reported event is currently underway.Device pending return.
 
Event Description
It was reported that during a procedure, air embolism was allegedly reported.The current status of the patient is unknown.
 
Event Description
It was reported that during a procedure, air embolism was allegedly reported.The current status of the patient is unknown.
 
Manufacturer Narrative
H10: manufacturing review: a lot history review, a device history record review and complete manufacturing review could not be conducted for the investigation as the lot number is unknown.Investigation summary: one introducer needle was returned for evaluation.Visual, microscopic visual and functional evaluations were performed on the returned device.The investigation is confirmed for the reported air leak as multiple cracks were observed on the proximal hub of the introducer needle.Also, leaks were observed from the needle upon infusion of water, only air was aspirated back into the syringe.A definitive root cause could not be determined based upon the available information.Labeling review: a review of product labeling documentation (e.G.Procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H10: g3, h6 (device: a1415) h11: h6 (device, method, result and conclusion) h11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : see h10.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POWERPORT ISP M.R.I. IMPLANTABLE PORT, GROSHONG SINGLE-LUMEN, 8F
Type of Device
PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
MDR Report Key11129916
MDR Text Key225480306
Report Number3006260740-2020-21206
Device Sequence Number1
Product Code LJT
UDI-Device Identifier00801741027512
UDI-Public(01)00801741027512
Combination Product (y/n)N
PMA/PMN Number
K063377
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 02/16/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number9808560
Device Catalogue Number9808560
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/08/2021
Initial Date Manufacturer Received 12/14/2020
Initial Date FDA Received01/07/2021
Supplement Dates Manufacturer Received07/07/2021
Supplement Dates FDA Received07/12/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Life Threatening;
-
-