Device evaluated by manufacturer: the device was returned for analysis: a visual examination identified that the balloon folds were tightly wrapped.No issues were noted with the balloon of the device that may have potentially contributed to the complaint incident.The markerbands and blades and tip section of the device were visually and microscopically examined, and no issues were noted with the markerbands, blades or tip of the device that may have potentially contributed to the complaint incident.All blades were present and fully bonded to the balloon material.A visual and tactile examination found that the device was received in two sections as a result of a break in the hypotube.The break was located at 20.9 cm distal to the strain relief.Both section of the hypotube were kinked near the break site.This type of damage is consistent with excessive force that could have been applied to the delivery system.A visual and tactile examination found a rotational twisting of the outer.This twisting stretched from the proximal balloon sleeve and extended proximally for a total length of 1cm.No damage was observed to the inner/wire lumen at the same location.This type of damage is consistent with excessive force that could have been applied to the delivery system.
|
Reportable based on device analysis completed on (b)(6) 2020.It was reported that the device was unable to cross the lesion.Vascular access was obtained via right femoral artery.The 60% stenosed target lesion area, in-stent restenosis with a >45 degrees bend was located in a moderately tortuous and moderately calcified left anterior descending artery.A 10mmx3.00mm wolverine coronary cutting balloon was selected for use in a percutaneous coronary intervention.During the procedure, the balloon was delivered with a non-bsc support wire.However, the device did not cross the lesion due to severe calcification.The device was simply removed from the patient's body.Another wolverine was used and crossed the lesion and completed the procedure.There were no complications reported and the patient was stable.However, device investigation revealed a hypotube break located at 20.9 cm distal to the strain relief.
|