Model Number 7122/65
Device Problems Failure to Capture (1081); Failure to Sense (1559)
Patient Problems No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/18/2020
Event Type  Injury  
Manufacturer Narrative
This product is registered as a combination product.The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Event Description
It was reported the patient presented in the clinic for a follow-up.Upon interrogation, it was observed the patient's right ventricular (rv) lead exhibited failure to sense r-waves and failure to capture.The rv lead was explanted and replaced on (b)(6) 2020.The patient was in stable condition.
 
Manufacturer Narrative
The reported event of failure to sense and failure to capture were not confirmed.As received, a partial lead was returned for analysis.Electrical testing that could be performed did not find any indication of conductor fractures or internal shorts.Visual and x-ray inspection of the lead did not find any anomalies with the exception of procedural damage.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DURATA STS OPTIM ACTIVE FIXATION
Type of Device
DEFIBRILLATION LEAD
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
sonali arangil
15900 valley view court
sylmar, CA 91342
MDR Report Key11133625
MDR Text Key225596085
Report Number2017865-2020-23644
Device Sequence Number1
Product Code LWS
UDI-Device Identifier05414734502153
UDI-Public05414734502153
Combination Product (y/n)N
Reporter Country CodeIN
PMA/PMN Number
P950022
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 01/25/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/30/2023
Device Model Number7122/65
Device Catalogue Number7122-65
Device Lot NumberA000097254
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/06/2021
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/19/2020
Initial Date FDA Received01/07/2021
Supplement Dates Manufacturer Received01/19/2022
Supplement Dates FDA Received01/25/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/20/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-