Description of issue: the patient reported that she felt itching.The ice pack was found out to be leaking on the long side of the package (not near a seam).Clinical staff immediately flushed the patients eye at an eyewash and notified the treating physicians on-call pager.A physician came to bedside and directed clinical staff to continue flushing the patients eye for 5-10 mins.The patients eye was flushed for the complete time.On follow-up thereafter, no effects were noted.Follow-up: on review with the staff member who used the ice pack, they had squeezed the packaging to open it.However, the use instructions indicate that it is to be folded over and shaken.There are also pictures on the front of the ice pack.Staff member was re-educated on use of the item and need to confirm packaging is intact before patient use.
|