This report was initially submitted following notification from a customer in the united states regarding a misidentification of clostridium perfringens as bacillus clausii when using the vitek® ms instrument (ref.(b)(4), serial # (b)(6).According to the vilink alert tool criteria, a fine-tuning was needed during the tests made on (b)(6) 2020.The customer¿s calibrator¿s spot preparation quality was not optimal.The calibrator ¿all peaks¿ values were quite heterogeneous.This could be explained by a non-optimal spot preparation of the calibrator strain.The expected identification is clostridium perfringens which is present in vitek® ms v3.2 knowledge base.The analysis of the mzml samples files showed the identification was obtained with a low identification score (-0.35) which is also near the acceptable limit for giving an ¿identification¿ result or a ¿no identification¿ result (-0.40).The root cause of the customer¿s results was determined to be due to non-optimal spot preparation, and fine-tuning.Recommended to check sample preparation with the customer, especially the quality of the strains used (age of the strain, culture media used) and the operators skills regarding spot preparation.See section h10.
|