MEDTRONIC PUERTO RICO OPERATIONS CO. ACTIVA; IMPLANTED SUBCORTICAL ELECTRICAL STIMULATOR (MOTOR DISORDERS) Back to Search Results
Model Number 37612
Device Problems Failure to Interrogate (1332); Improper or Incorrect Procedure or Method (2017); Battery Problem (2885); Communication or Transmission Problem (2896)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
Date of event: the event date is unknown.Common device name: please note that this device was used in an off-label manner as it was implanted for dystonia.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported the patient might have gotten into three overdischarges.The physician was unable to interrogate the patient's device with the tablet and the recharger is unable to connect to the patient's device.They were not with the patient.It was suggested the physician call if they need further assistance.No patient symptoms were reported.
 
Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
The overdischarge was confirmed.The cause of the connection issue was due to the over discharge and the device not being charged above 25%.They were able to connect to the ins once it was charged above 50%.The issue was resolved.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACTIVA
Type of Device
IMPLANTED SUBCORTICAL ELECTRICAL STIMULATOR (MOTOR DISORDERS)
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
MDR Report Key11140258
MDR Text Key225782712
Report Number3004209178-2021-00415
Device Sequence Number1
Product Code MRU
UDI-Device Identifier00643169529762
UDI-Public00643169529762
Combination Product (y/n)N
PMA/PMN Number
H020007
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 01/12/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/28/2018
Device Model Number37612
Device Catalogue Number37612
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 01/07/2021
Initial Date FDA Received01/08/2021
Supplement Dates Manufacturer Received01/08/2021
Supplement Dates FDA Received01/12/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-