Model Number 5826 |
Device Problem
Pacing Problem (1439)
|
Patient Problems
No Code Available (3191); Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
|
Event Date 12/18/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
|
|
Event Description
|
It was reported that the patient presented for generator change due to eri.The patient's device was interrogated prior to the procedure and the patient was 100% pacemaker dependent with no escape rhythm.When the physician proceeded to cut skin with electrocautery, the device exhibited common-mode inhibition behavior and the patient's rhythm became asystole for approximately 3 minutes until the device started to recover and resumed pacing again.The physician acknowledged that the device did go into common-mode inhibition behavior and they should have had an extra torque wrench and pacing cables connected to the analyzer ready to pace to anticipate for the asystole.The device was replaced.The patient was stable prior to the surgery, nauseous and threw up immediately post-surgery, and stable once transferred to the post-op recovery unit.
|
|
Manufacturer Narrative
|
The reported of pacing problem was not confirmed.Visual examination noted burn marks on device can from exposure to electrosurgical cautery.User¿s manual indicated that electrosurgical cautery can inhibit the pulse generator operation.As received the device operating at normal mode.The results of electrical, stress/environmental tests performed indicate the pacer is exhibiting normal device characteristics.Longevity assessment was performed based on field settings and device was in the normal range of operation with appropriate remaining longevity.Device image was successfully saved.
|
|
Search Alerts/Recalls
|