• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LIVANOVA USA, INC. PULSE GEN MODEL 106; GENERATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LIVANOVA USA, INC. PULSE GEN MODEL 106; GENERATOR Back to Search Results
Model Number 106
Device Problem Premature End-of-Life Indicator (1480)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/16/2020
Event Type  malfunction  
Event Description
It was reported that a patient implanted in may is showing 50% battery life and the physician is trying to determine if the battery is depleting normally or if it could be depleting prematurely.No additional relevant information has been received to date.
 
Event Description
The 25% indicator was displayed due to the effect of the high settings and high duty cycle on the battery voltage, and isn¿t indicative that the battery is depleting faster than expected.In actuality, based on the data from the last office visit it is more likely that close to 50% of battery capacity remains.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PULSE GEN MODEL 106
Type of Device
GENERATOR
Manufacturer (Section D)
LIVANOVA USA, INC.
100 cyberonics blvd
houston TX 77058
MDR Report Key11141395
MDR Text Key225826289
Report Number1644487-2021-00046
Device Sequence Number1
Product Code LYJ
Combination Product (y/n)N
PMA/PMN Number
P970003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 02/15/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date01/19/2022
Device Model Number106
Device Lot Number205179
Was Device Available for Evaluation? No
Event Location Other
Initial Date Manufacturer Received 12/16/2020
Initial Date FDA Received01/08/2021
Supplement Dates Manufacturer Received01/21/2021
Supplement Dates FDA Received02/15/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age3 YR
-
-