It was reported that during use with patient of this disposable pressure transducer (dpt) , central venous pressure (cvp) value was 25mmhg, which was inaccurate.Originally the patient was being monitored with a dpt that showed cvp value of 6-7mmhg.When this dpt was changed for a new one the value raised to 25mmhg without transfusion nor injection from the cvc catheter lumens.The expected value according to the patient's clinical condition was around 10mmhg so the user did not expect to see a value of 25mmhg.They changed to a third dpt and the value dropped to 13mmhg immediately.Therefore this complaint is for the second transducer that gave a value of 25mmhg.There were no error messages.There were no other inaccurate values apart from cvp.There was no allegation of patient injury.The product was available for evaluation.Patient demographics are not available.
|
One single dpt kit was received for evaluation.The reported event of inaccurate readings was not confirmed.The dpt zeroed and sensed pressure accurately on pressure monitor.Pressure did not show any drift during output drift testing and met specification.Electrical testing showed that both input impedance and output impedance were within specifications.Zero-offset also met specification per drawing 270378 rev.T.No leakage or occlusion was detected from the kit during pressure test.No visible damage was observed from the kit.Poor dynamic response can be caused by air bubbles, clotting, and excessive lengths of tubing, excessively compliant pressure tubing, small bore tubing, loose connections, or leaks.The assembly may be tested for dynamic response by observing the pressure waveform on an oscilloscope or monitor.Bedside determination of the dynamic response of the catheter, monitor, kit and transducer system is done after the system is flushed, attached to the patient, zeroed and calibrated.A square-wave test may be performed by pulling the snap tab device and releasing quickly.Pressure readings can change quickly and dramatically because of loss of proper calibration, loose connection, or air in the system.Pressure readings should correlate with the patient¿s clinical manifestations.It is not known if user or procedural factors may have contributed to the stated event.In this event, there was no patient compromise noted.Complaint histories for all reported events are reviewed against trending control limits on a monthly basis and any excursions above the control limits are assessed and documented as a part of the monthly review.
|