BOSTON SCIENTIFIC CORPORATION TRAPEZOID RX; LITHOTRIPTOR, BILIARY MECHANICAL Back to Search Results
Model Number M00510860
Device Problem Break (1069)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/16/2020
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The device has not been received for analysis.Upon receipt and completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental mdr will be filed.
 
Event Description
It was reported to boston scientific corporation that a trapezoid rx lithotripter basket was used in the common bile duct during an endoscopic retrograde cholangiopancreatography (ercp) procedure performed on (b)(6) 2020.According to the complainant, during the procedure, the basket was noted to be difficult to open.The technician applied force to the handle in order to open the basket but the handle cannula broke.There was no stone inside the basket when the handle cannula broke.The procedure was completed with another trapezoid rx basket.There were no patient complications reported as a result of this event.The patient's condition following the procedure was reported to be good.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRAPEZOID RX
Type of Device
LITHOTRIPTOR, BILIARY MECHANICAL
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
780 brookside drive
spencer IN 47460
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key11143178
MDR Text Key225931768
Report Number3005099803-2020-06679
Device Sequence Number1
Product Code LQC
UDI-Device Identifier08714729296379
UDI-Public08714729296379
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K040447
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 01/08/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/07/2021
Device Model NumberM00510860
Device Catalogue Number1086
Device Lot Number0025498481
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/16/2020
Initial Date FDA Received01/08/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/07/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-