Model Number OER-PRO
Device Problem Particulates (1451)
Patient Problem No Patient Involvement (2645)
Event Date 12/15/2020
Event Type  malfunction  
Manufacturer Narrative
An olympus field service engineer (fse) was dispatched to confirm the reported issue.The fse serviced the entire equipment, replaced the necessary hoses and disinfected according to specifications.No other issues were reported.If additional information is obtained a supplemental report will be filed.
 
Event Description
A user facility reported black specs on an endoscope reprocessor.There was no patient involvement or injuries reported.No additional information has been obtained.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's investigation and to correct information provided on the initial report.The following sections were updated and/or corrected.Based on the legal manufacturer's investigation and the internal risk summary tool, this supplemental report is to inform that upon further review, this is not a reportable malfunction or a potential adverse event.Per the legal manufacturer, there is no potential for this issue to cause or contribute to death or serious injury if the malfunction were to recur.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOSCOPE REPROCESSOR
Type of Device
ENDOSCOPE REPROCESSOR
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key11148939
MDR Text Key238137174
Report Number8010047-2021-01175
Device Sequence Number1
Product Code FEB
UDI-Device Identifier04953170258589
UDI-Public04953170258589
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K103264
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 03/02/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberOER-PRO
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/15/2020
Initial Date FDA Received01/11/2021
Supplement Dates Manufacturer Received03/01/2021
Supplement Dates FDA Received03/02/2021
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-