MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 BARD INLAY OPTIMA URETERAL STENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 BARD INLAY OPTIMA URETERAL STENT Back to Search Results
Model Number 788426
Device Problem Material Fragmentation (1261)
Patient Problems No Patient Involvement (2645); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/01/2020
Event Type  malfunction  
Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
 
Event Description
It was reported that the pigtail end of the ureteral stent was found ruptured after the package was opened.
 
Manufacturer Narrative
The reported event was inconclusive as no sample returned for evaluation.It was unknown whether the device had met specifications.The product used for the treatment purposes.It was unknown whether the product had caused the reported failure.A potential root cause for this failure mode could be due to material incompatibility.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The labeling review was not performed due to the labeling could not have prevented the reported failure.H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : the device was not returned.
 
Event Description
It was reported that the pigtail end of the ureteral stent was found ruptured upon opening the package.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARD INLAY OPTIMA URETERAL STENT
Type of Device
URETERAL STENT
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
MDR Report Key11150640
MDR Text Key226278714
Report Number1018233-2020-22433
Device Sequence Number1
Product Code FAD
UDI-Device Identifier00801741015687
UDI-Public(01)00801741015687
Combination Product (y/n)N
PMA/PMN Number
K043193
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,other
Type of Report Initial,Followup
Report Date 04/04/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/11/2024
Device Model Number788426
Device Catalogue Number788426
Device Lot NumberNGDV1980
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 12/19/2020
Initial Date FDA Received01/11/2021
Supplement Dates Manufacturer Received04/04/2021
Supplement Dates FDA Received04/26/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
-
-