Information was received from a healthcare provider via a manufacturer representative regarding a patient who was implanted with an implantable neurostimulator (ins) for gastric stimulation, gastrointestinal/pelvic floor.It was reported that the patient had reported increased nausea and vomiting.It was found that the impedance was greater than 20,000 ohms during an impedance check.The doctor planned to do a revision on (b)(6) 2021.The issue was not resolved at the time of the report.
|
Additional information was received from a manufacturer representative (rep).The rep confirmed that the health care professional (hcp) did the revision on (b)(6) 2021, and that the hcp removed the leads from the battery, wiped them off, reinserted the leads and tightened the set screws.The impedance then returned to normal at 490.The rep stated they obtained this information from the or scheduler.
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|