Model Number ALLURA XPER FD20 |
Device Problem
Fluid/Blood Leak (1250)
|
Patient Problems
No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/15/2020 |
Event Type
malfunction
|
Event Description
|
It has been reported to philips that during a procedure oil leaked from the c-arm.The procedure was aborted and the patient moved to another room.No harm to the patient and user has been reported to philips.Philips has started an investigation of this complaint.
|
|
Manufacturer Narrative
|
A philips engineer inspected the system onsite and identified that there was a small hole in the (x-ray tube) cooling hose causing the oil leak.The philips engineer replaced the cooling hose and the system was returned to use in good working order.
|
|