Model Number CDS0701-XTW
Device Problems Difficult to Open or Close (2921); Noise, Audible (3273); Physical Resistance/Sticking (4012)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/31/2020
Event Type  malfunction  
Manufacturer Narrative
The clip remains in patient.The device will not be returned for evaluation.Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
 
Event Description
This is filed to report clip jumping.It was reported that this was a mitraclip procedure to treat degenerative mitral regurgitation (mr) with a grade of 4.The clip delivery system (cds) was advanced to the left atrium; however, during the initial clip opening, there was resistance turning the arm positioner and a squeaky noise was heard.Then the clip jumped open.Therefore, the clip was re-locked and closed.The clip then opened as intended and the procedure continued as normal.The clip was deployed.One clip was implanted, reducing mr to 1.There were no adverse patient effects and no clinically significant delay in the procedure.No additional information was provided.
 
Manufacturer Narrative
The device was not returned.A review of the lot history record revealed no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no similar complaints from the lot.All available information was investigated and without the device to analyze, a cause for the reported clip jumpiness and resistance on the arm positioner resulting in noise could not be determined.There is no indication of a product issue with respect to manufacture, design, or labeling.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MITRACLIP G4 CLIP DELIVERY SYSTEM
Type of Device
MITRAL VALVE REPAIR DEVICES
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
MDR Report Key11164365
MDR Text Key226623622
Report Number2024168-2021-00403
Device Sequence Number1
Product Code NKM
UDI-Device Identifier08717648231001
UDI-Public08717648231001
Combination Product (y/n)N
PMA/PMN Number
P100009
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 02/23/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/16/2021
Device Model NumberCDS0701-XTW
Device Catalogue NumberCDS0701-XTW
Device Lot Number00917U191
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 12/31/2020
Initial Date FDA Received01/13/2021
Supplement Dates Manufacturer Received02/02/2021
Supplement Dates FDA Received02/23/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
STEERABLE GUIDE CATHETER; STEERABLE GUIDE CATHETER
Patient Age71 YR
-
-