The product was not returned for evaluation.Without the return of the device, the root cause of the problem cannot be determined.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
|
The patient was undergoing a thrombectomy procedure in the superficial femoral artery (sfa) using an indigo system aspiration catheter 8 (cat8) and a non-penumbra sheath.It should be noted that the patient had a tortuous aorta-iliac bifurcation.During the procedure, after several passes were completed using the cat8, the physician lost suction while torquing the cat8 and upon removal, realized it was kinked mid-shaft.Therefore, the cat8 was removed and a new cat8 was used to continue the procedure.Next, after a few passes were completed with the second cat8, the physician lost suction after torquing the cat8 on the final pass.Upon removal, the cat8 broke into two pieces.It was reported that the hub of the cat8 was pulled out and the distal half of the catheter was contained in the sheath.Therefore, this sheath was also removed.The procedure was completed at this point and no additional passes were needed.There was no report of an adverse effect to the patient.
|