MAUDE Adverse Event Report: ST. JUDE MEDICAL ADVISOR HD GRID MAPPING CATHETER, SENSOR ENABLED; CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL ADVISOR HD GRID MAPPING CATHETER, SENSOR ENABLED; CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING Back to Search Results
Model Number D-AVHD-DF16
Device Problems Signal Artifact/Noise (1036); Display or Visual Feedback Problem (1184)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/05/2020
Event Type  malfunction  
Manufacturer Narrative
One bi-directional, curve d-f, sensor enabled, advisor hd grid mapping catheter was received for evaluation.A request for additional analysis was made for an advisor hd mapping catheter with a complaint mode related to electrodes 3-4 having no signal and noise.Visual inspection concluded that electrodes 3-4 had been displaced.Functional testing revealed an open circuit of electrodes 3-4 which were isolated to the weld joint of the electrode ring and the wire within the paddle.Device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The root cause of the fractured conductor wire and displaced and flared electrodes is consistent with damage during use.
 
Event Description
During the procedure, there was distortion on the a spline of the hd grid catheter on the mapping system.There was also noise on the a spline egm signals on the recording system.The cables for the grid catheter were verified to be pinned correctly.The cables were replaced to a new set of cables and pinned correctly again with no resolution.The hd grid catheter was replaced, and the procedure was completed with no adverse patient consequences.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVISOR HD GRID MAPPING CATHETER, SENSOR ENABLED
Type of Device
CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING
Manufacturer (Section D)
ST. JUDE MEDICAL
5050 nathan lane north
plymouth MN 55442
Manufacturer (Section G)
ST. JUDE MEDICAL
5050 nathan lane north
plymouth MN 55442
Manufacturer Contact
stephanie o' sullivan
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key11169506
MDR Text Key226793201
Report Number3005334138-2021-00017
Device Sequence Number1
Product Code DRF
UDI-Device Identifier05415067028198
UDI-Public05415067028198
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K172393
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 01/14/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/31/2023
Device Model NumberD-AVHD-DF16
Device Catalogue NumberD-AVHD-DF16
Device Lot Number7619862
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/16/2020
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 01/08/2021
Initial Date FDA Received01/14/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/11/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-