• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT VASCULAR MITRACLIP G4 CLIP DELIVERY SYSTEM; MITRAL VALVE REPAIR DEVICES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR MITRACLIP G4 CLIP DELIVERY SYSTEM; MITRAL VALVE REPAIR DEVICES Back to Search Results
Model Number CDS0701-XTW
Device Problem Difficult to Open or Close (2921)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/04/2021
Event Type  malfunction  
Manufacturer Narrative
The device is expected to be returned for investigation.It has not yet been received.A follow-up report will be submitted with all additional relevant information.
 
Event Description
This is being filed to report gripper actuation issue.It was reported that this was a mitraclip procedure to treat a degenerative mitral regurgitation (mr) with grade of 4.The clip delivery system (cds) was advanced to the mitral valve and grasping was performed.However, one of the gripper arms did not go down.Troubleshooting was performed such as cycling the lock lever and re-closing the clip but the gripper arm did not come down.Therefore, the physician decided to invert the clip and remove the cds.A new cds was used to complete the procedure.One clip was implanted, reducing mr to 1.There was no adverse patient effect and no clinically significant delay in the procedure.No additional information was provided.
 
Manufacturer Narrative
The returned device analysis could not confirm the reported difficult to open or close (gripper actuation) as both grippers were able to lower and raise as intended without issue.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot which could have contributed to the reported issues.Additionally, a review of the complaint history did not identify a lot specific issue.Based on the information reviewed, a cause for the reported difficult to open or close (gripper actuation) could not be determined.There is no indication of a product issue with respect to manufacture, design or labeling.Na.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MITRACLIP G4 CLIP DELIVERY SYSTEM
Type of Device
MITRAL VALVE REPAIR DEVICES
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
MDR Report Key11177564
MDR Text Key227045128
Report Number2024168-2021-00478
Device Sequence Number1
Product Code NKM
UDI-Device Identifier08717648231001
UDI-Public08717648231001
Combination Product (y/n)N
PMA/PMN Number
P100009
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 03/01/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/20/2021
Device Model NumberCDS0701-XTW
Device Catalogue NumberCDS0701-XTW
Device Lot Number00921U123
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/20/2021
Initial Date Manufacturer Received 01/04/2021
Initial Date FDA Received01/15/2021
Supplement Dates Manufacturer Received02/04/2021
Supplement Dates FDA Received03/01/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-