Model Number CDS0701-XTW |
Device Problem
Difficult to Open or Close (2921)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 01/04/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
The device is expected to be returned for investigation.It has not yet been received.A follow-up report will be submitted with all additional relevant information.
|
|
Event Description
|
This is being filed to report gripper actuation issue.It was reported that this was a mitraclip procedure to treat a degenerative mitral regurgitation (mr) with grade of 4.The clip delivery system (cds) was advanced to the mitral valve and grasping was performed.However, one of the gripper arms did not go down.Troubleshooting was performed such as cycling the lock lever and re-closing the clip but the gripper arm did not come down.Therefore, the physician decided to invert the clip and remove the cds.A new cds was used to complete the procedure.One clip was implanted, reducing mr to 1.There was no adverse patient effect and no clinically significant delay in the procedure.No additional information was provided.
|
|
Manufacturer Narrative
|
The returned device analysis could not confirm the reported difficult to open or close (gripper actuation) as both grippers were able to lower and raise as intended without issue.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot which could have contributed to the reported issues.Additionally, a review of the complaint history did not identify a lot specific issue.Based on the information reviewed, a cause for the reported difficult to open or close (gripper actuation) could not be determined.There is no indication of a product issue with respect to manufacture, design or labeling.Na.
|
|
Search Alerts/Recalls
|