• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT GMBH MAGNESIUM; PHOTOMETRIC METHOD, MAGNESIUM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT GMBH MAGNESIUM; PHOTOMETRIC METHOD, MAGNESIUM Back to Search Results
Model Number 03P6832
Device Problem High Test Results (2457)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/28/2020
Event Type  malfunction  
Manufacturer Narrative
No additional patient information is available.A review of tickets was performed for lot number 01965un20.The ticket search determined that there is normal complaint activity for the reagent lot.A review of tracking and trending did not identify any trends for the complaint issue.Return testing was not completed as returns were not available.The historical performance of reagent lot 01965un20 was evaluated using field data.The patient data was analyzed and compared to an established control limit.This evaluation indicated that the patient median result for lot 01965un20 is within the established control limits.A review of the manufacturing documentation did not identify any issues associated with the complaint issue.A review of labeling concluded that the issue is sufficiently addressed.Based on the investigation no systemic issue or deficiency of the magnesium assay for lot 01965un20 was identified.
 
Event Description
The customer generated falsely elevated magnesium results while using the architect c16000 processing module.The customer's normal range is 1.6 - 2.6 mg/dl.The following information was provided: sid (b)(6) nitial result: 7.2 mg/dl, repeat 1.9 mg/dl on a different analyzer.Initial result: 7.2 mg/dl, repeat 1.9 mg/dl on a different analyzer.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAGNESIUM
Type of Device
PHOTOMETRIC METHOD, MAGNESIUM
Manufacturer (Section D)
ABBOTT GMBH
max-planck-ring 2
wiesbaden 65205
GM  65205
Manufacturer (Section G)
ABBOTT GMBH
max-planck-ring 2
wiesbaden 65205
GM   65205
Manufacturer Contact
christian lee
100 abbott park road
dept. 09b9, lc cp01-3
abbott park, IL 60064-3537
2246682940
MDR Report Key11181645
MDR Text Key240406133
Report Number3002809144-2021-00033
Device Sequence Number1
Product Code JGJ
UDI-Device Identifier00380740161637
UDI-Public00380740161637
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K181748
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 01/15/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/11/2021
Device Model Number03P6832
Device Catalogue Number03P68-32
Device Lot Number01965UN20
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/28/2020
Initial Date FDA Received01/15/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/09/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-