• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT VASCULAR MITRACLIP G4 CLIP DELIVERY SYSTEM; MITRAL VALVE REPAIR DEVICES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR MITRACLIP G4 CLIP DELIVERY SYSTEM; MITRAL VALVE REPAIR DEVICES Back to Search Results
Model Number CDS0701-NTW
Device Problems Incomplete Coaptation (2507); Unintended Movement (3026)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/05/2021
Event Type  Injury  
Manufacturer Narrative
The clip remains in patient.The device will not be returned for evaluation.Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.Na.
 
Event Description
This is filed to report unintended movement, single leaflet device attachment/slda and medical intervention.It was reported that this was a mitraclip procedure to treat degenerative mitral regurgitation (mr) with a grade of 4+.It was noted a hyper mobile anterior leaflet.The clip delivery system (cds) was advanced to the mitral valve, and the clip was placed.However, upon release, tension was felt with the clip.The clip torqued to the anterior direction which caused the clip to become detached from the posterior leaflet but remain attached to the anterior leaflet (single leaflet device attachment/slda).Two additional clips were deployed on the lateral and medial side of the clip to stabilize the slda.Three clips were implanted, reducing mr to 2.There were no adverse patient effects and no clinically significant delay in the procedure.No additional information was provided.
 
Manufacturer Narrative
The device was not returned for analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have resulted in this event.Additionally, a review of the complaint history identified no other similar complaint reported from this lot.All available information was investigated, and a conclusive cause for unintended movement could not be determined in this complaint.The reported single leaflet device attachment/slda was a cascading effect of the unintended movement due to the torqueing/tension on clip causing posterior leaflet detachment.The reported additional therapy/non-surgical treatment appears to be due to case specific circumstances.There is no indication of a product issue with respect to manufacture, design or labeling.Na.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MITRACLIP G4 CLIP DELIVERY SYSTEM
Type of Device
MITRAL VALVE REPAIR DEVICES
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
MDR Report Key11186174
MDR Text Key227572678
Report Number2024168-2021-00511
Device Sequence Number1
Product Code NKM
UDI-Device Identifier08717648230967
UDI-Public08717648230967
Combination Product (y/n)N
PMA/PMN Number
P100009
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 02/24/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/19/2021
Device Model NumberCDS0701-NTW
Device Catalogue NumberCDS0701-NTW
Device Lot Number00820U110
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 01/05/2021
Initial Date FDA Received01/18/2021
Supplement Dates Manufacturer Received02/03/2021
Supplement Dates FDA Received02/24/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
STEERABLE GUIDE CATHETER
Patient Outcome(s) Required Intervention;
Patient Age71 YR
-
-