• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT VASCULAR MITRACLIP G4 CLIP DELIVERY SYSTEM; MITRAL VALVE REPAIR DEVICES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR MITRACLIP G4 CLIP DELIVERY SYSTEM; MITRAL VALVE REPAIR DEVICES Back to Search Results
Model Number CDS0701-NT
Device Problem Difficult to Open or Close (2921)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/05/2021
Event Type  malfunction  
Manufacturer Narrative
The device will not be returned for evaluation, the device was reportedly discarded.Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
 
Event Description
This is being filed to report the gripper actuation issue.It was reported that this was a mitraclip procedure to treat degenerative mitral regurgitation (mr) with a grade of 4.One clip was implanted without issue.A second clip delivery system (cds) was advanced to the mitral valve and the leaflets grasped however one of the grippers did not lower.Troubleshooting was performed however unsuccessful therefore the cds was removed.Two additional clips were implanted, reducing mr to 2.There was no clinically significant delay in the procedure and no adverse patient effects.No additional information was provided.
 
Manufacturer Narrative
The device was not returned for analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot.Additionally, a review of the complaint history did not identify any similar incidents.Based on the information reviewed and without the device to analyze, a cause for the reported difficult to open or close (gripper actuation) could not be determined.There is no indication of a product issue with respect to manufacture, design or labeling.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MITRACLIP G4 CLIP DELIVERY SYSTEM
Type of Device
MITRAL VALVE REPAIR DEVICES
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
MDR Report Key11186292
MDR Text Key230276512
Report Number2024168-2021-00515
Device Sequence Number1
Product Code NKM
UDI-Device Identifier08717648230943
UDI-Public08717648230943
Combination Product (y/n)N
PMA/PMN Number
P100009
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 01/26/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/03/2021
Device Model NumberCDS0701-NT
Device Catalogue NumberCDS0701-NT
Device Lot Number00803U250
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 01/05/2021
Initial Date FDA Received01/18/2021
Supplement Dates Manufacturer Received01/25/2021
Supplement Dates FDA Received01/26/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
1 IMPLANTED MITRACLIP; STEERABLE GUIDE CATHETER
Patient Age87 YR
Patient Weight46
-
-