Model Number CD2357-40Q |
Device Problems
Premature Discharge of Battery (1057); Battery Problem (2885)
|
Patient Problems
No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/11/2021 |
Event Type
Injury
|
Manufacturer Narrative
|
The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.The device is included in the battery performance alert advisory issued by abbott on 28 august 2017.
|
|
Event Description
|
Following the battery performance alert (bpa) advisory, a bpa was received by the clinician and awaiting explant.Further information was requested, but not yet available.
|
|
Event Description
|
New information notes that following the battery performance alert (bpa) advisory, a bpa was received by the clinician and the device was explanted.Patient was stable.
|
|
Manufacturer Narrative
|
Premature battery depletion was confirmed by analysis.No sources of high current were noted.The cause of the premature battery depletion was consistent with li cluster formation.From these analyses, in the absence of high current draw, it is probable that the premature battery depletion was caused by a lithium cluster induced short circuit.Li clusters are a known depletion mechanism for these advisory products that has been investigated and associated with a field action in october 2016.
|
|