• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT VASCULAR MITRACLIP G4 CLIP DELIVERY SYSTEM; MITRAL VALVE REPAIR DEVICES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR MITRACLIP G4 CLIP DELIVERY SYSTEM; MITRAL VALVE REPAIR DEVICES Back to Search Results
Model Number CDS0701-NTW
Device Problem Incomplete Coaptation (2507)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/04/2021
Event Type  malfunction  
Manufacturer Narrative
The clip remains in patient.The device will not be returned for evaluation.Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.Na.
 
Event Description
This is filed for single leaflet device attachment (slda).It was reported that this was a mitraclip procedure to treat mixed mitral regurgitation (mr) with a grade of 4.The posterior leaflet was noted to be shorter, restricted, thickened and calcified.One clip was implanted without issue.A second clip was implanted; however, after deployment, a single leaflet device attachment (slda) occurred, with the clip detaching from the posterior leaflet.The clip remained attached to the anterior leaflet and appeared stable on the remaining leaflet.No additional intervention was performed and mr was reduced to grade 2.There was no adverse patient effect or a clinically significant delay in the procedure.No additional information was provided.
 
Manufacturer Narrative
The device was not returned for analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.A review of the complaint history identified no similar incidents reported form this lot.Based on the available information, a cause for the reported single leaflet device attachment (slda) could not be determined.There is no indication of a product issue with respect to manufacture, design, or labeling.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MITRACLIP G4 CLIP DELIVERY SYSTEM
Type of Device
MITRAL VALVE REPAIR DEVICES
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
MDR Report Key11189906
MDR Text Key227571941
Report Number2024168-2021-00546
Device Sequence Number1
Product Code NKM
UDI-Device Identifier08717648230967
UDI-Public08717648230967
Combination Product (y/n)N
PMA/PMN Number
P100009
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 02/01/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/21/2021
Device Model NumberCDS0701-NTW
Device Catalogue NumberCDS0701-NTW
Device Lot Number00821U234
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 01/04/2021
Initial Date FDA Received01/18/2021
Supplement Dates Manufacturer Received01/25/2021
Supplement Dates FDA Received02/01/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
OTHER IMPLANTED CLIP
Patient Age81 YR
Patient Weight84
-
-