SYNTHES GMBH SAW BLADE 22 X 8.0 X 0.38MM F/SAG SAW; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT
|
Back to Search Results |
|
Catalog Number 03.000.303S |
Device Problems
Vibration (1674); Dull, Blunt (2407); Device-Device Incompatibility (2919)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/10/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.As of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.Concomitant med products and therapy dates: electric pen drive device; (b)(6) 2021.Initial reporter phone: not available.Udi: (b)(4).
|
|
Event Description
|
It was reported from (b)(6) that during an unspecified surgical procedure, it was observed that the saw blade cutter device was not sharp enough and was shaky (would not attach fully) when inserted into electric pen drive device.There were no reports of delays in the surgical procedure.It was not reported if a spare device was available for use.There was patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
|
|
Manufacturer Narrative
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: this device was returned for evaluation; however, during pre-repair assessment, it was determined that the device passed all operational specifications and no failure was identified.Therefore, the reported condition was not confirmed and an assignable root cause was not determined.A device history review was performed, and no non-conformances were detected related to the reported condition.
|
|