• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WRIGHT MEDICAL TECHNOLOGY, INC. INFINITY TOTAL ANKLE SYSTEM; PROSTHESIS, ANKLE, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WRIGHT MEDICAL TECHNOLOGY, INC. INFINITY TOTAL ANKLE SYSTEM; PROSTHESIS, ANKLE, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER Back to Search Results
Device Problem Patient Device Interaction Problem (4001)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/23/2020
Event Type  Injury  
Manufacturer Narrative
The device was not returned for evaluation.Two radiographic images were provided that depict infinity devices in-situ.It does appear to show some lucencies around the tibial tray pegs, but doesn't appear to have shifted; however, immediate postop images were not provided for comparison.
 
Event Description
It was reported that the patient underwent a total ankle replacement surgery.Allegedly, during a 3 year follow-up appointment, the patient's x-ray showed voids around the pegs of the talar implant.Physician believes that the implant may be loose.The physician does not believe it is an implant issue.The patient is a roofer and puts extreme load on the implants.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INFINITY TOTAL ANKLE SYSTEM
Type of Device
PROSTHESIS, ANKLE, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER
Manufacturer (Section D)
WRIGHT MEDICAL TECHNOLOGY, INC.
11576 memphis arlington rd
arlington TN 38002
Manufacturer (Section G)
WRIGHT MEDICAL TECHNOLOGY, INC.
11576 memphis arlington rd
arlington TN 38002
Manufacturer Contact
matthew parrish
1023 cherry rd
memphis, TN 38117
MDR Report Key11201354
MDR Text Key227750254
Report Number1043534-2021-00009
Device Sequence Number1
Product Code HSN
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial
Report Date 12/23/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 12/23/2020
Initial Date FDA Received01/20/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/06/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-