Model Number FGS-0635
Device Problem Failure to Transmit Record (1521)
Patient Problem Nausea (1970)
Event Date 12/31/2020
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
According to the reporter, the capsule failed to transmit after placement.The customer attempted pairing with two different recorders, but the capsule was not detected.Technical support instructed the customer to attempt pairing prior to placing capsule to ensure proper operation.The customer stated the patient experienced nausea, so another capsule was not placed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BRAVO
Type of Device
ELECTRODE, PH, STOMACH
Manufacturer (Section D)
GIVEN IMAGING LTD., YOQNEAM
yetsira 13 street
yoqneam 20692
IS  20692
Manufacturer (Section G)
GIVEN IMAGING LTD., YOQNEAM
yetsira 13 street
yoqneam 20692
IS   20692
Manufacturer Contact
amy beeman
161 cheshire lane, suite 100
plymouth, MN 55441
7632104064
MDR Report Key11201367
MDR Text Key227766412
Report Number9710107-2021-00044
Device Sequence Number1
Product Code FFT
UDI-Device Identifier07290101369707
UDI-Public07290101369707
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K102543
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Nurse
Type of Report Initial
Report Date 01/20/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/30/2022
Device Model NumberFGS-0635
Device Catalogue NumberFGS-0635
Device Lot Number51422F
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/31/2020
Initial Date FDA Received01/20/2021
Date Device Manufactured11/30/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-