ZIMMER SURGICAL, INC. ZMR AIR DERMATOME HANDPIECE Back to Search Results
Model Number N/A
Device Problems Loss of Power (1475); Output below Specifications (3004)
Patient Problems No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/08/2021
Event Type  malfunction  
Manufacturer Narrative
This event has been recorded by zimmer biomet under (b)(4).Once the investigation is completed, a follow-up/final report will be submitted.
 
Event Description
It was reported that the device lost power during surgery.No adverse events were reported as a result of this event.
 
Manufacturer Narrative
This event has been recorded by zimmer biomet under (b)(4).This medwatch is being filed to relay additional information.Review of the most recent repair record identified no repairs related to the reported event.Device is used for treatment.It was noted the device has not been returned regularly for recommended annual pm.A definitive root cause cannot be determined as no problem was found related to the reported event.The event cannot be confirmed.
 
Event Description
There is no additional information.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ZMR AIR DERMATOME HANDPIECE
Type of Device
DERMATOME
Manufacturer (Section D)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
MDR Report Key11201767
MDR Text Key227769897
Report Number0001526350-2021-00057
Device Sequence Number1
Product Code GFD
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 05/12/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number00880100100
Device Lot Number61919647
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/15/2021
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 01/15/2021
Initial Date FDA Received01/20/2021
Supplement Dates Manufacturer Received05/11/2021
Supplement Dates FDA Received05/12/2021
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberN/A
Patient Sequence Number1
-
-