Model Number 03R6501
Device Problem False Negative Result (1225)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/01/2021
Event Type  malfunction  
Manufacturer Narrative
An evaluation is in process.A follow-up report will be submitted when the evaluation is complete.No further patient information was provided.
 
Event Description
The customer obtained a (b)(6) result while using the alinity i processing module.The patient was a (b)(6) male undergoing chemotherapy for a malignant tumor.The sample generated non-reactive results of (b)(6) s/co on the alinity and positive results with the elisa and tppa assays.The customer further indicated (b)(6) 2021 the patient's sample was negative on roche luminescence 0.068 coi.On (b)(6) 2021 the patient's sample generated negative results on architect 0.26 s/co, rpr and elisa.No impact to patient management was reported.
 
Manufacturer Narrative
H6 health effect impact code: f26.H6 component code: g03001.D8 was this device serviced by a third party? no.The complaint investigation included a review of data and information provided by the customer, search for similar complaints, ticket trending review, labeling review, device history record review and in-house testing of the complaint lot.Trending review determined identified no trends associated with the complaint issue.Device history record review for the complaint lot did not identify any non-conformances or deviations.A retained reagent kit of the complaint lot was tested in a sensitivity setup.All specifications were met and no false non-reactive results were obtained, indicating that the sensitivity performance is not negatively impacted.Based on our investigation, no systemic issue or deficiency with the alinity i syphilis tp assay was identified.After further evaluation, the suspect medical device was changed from alinity i processing module, list number 03r65-01, serial (b)(6) throughout section d of this report to alinity i syphilis, list number 07p60-77 and lot 15228be01 from the same manufacturing site.Additionally section d11 was changed from the alinity i syphilis reagent to the alinity i processing module.
 
Manufacturer Narrative
The incorrect component code was entered in h6.The correct component code is g01003.No other corrections or updates are required.
 
Manufacturer Narrative
This follow up is submitted to populate fields d8 and/or h6 with data that had previously been provided in field h10.There is no change to the content of the data.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALINITY I SYPHILIS TP REAGENT KIT
Type of Device
ENZYME LINKED IMMUNOABSORPTION ASSAY, TREPONEMA PALLIDUM
Manufacturer (Section D)
ABBOTT GMBH
max-planck-ring 2
wiesbaden 65205
GM  65205
MDR Report Key11204965
MDR Text Key243602352
Report Number3002809144-2021-00042
Device Sequence Number1
Product Code LIP
UDI-Device Identifier00380740137366
UDI-Public00380740137366
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Type of Report Initial,Followup,Followup,Followup
Report Date 05/18/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/16/2021
Device Model Number03R6501
Device Catalogue Number07P60-77
Device Lot Number15228BE01
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 01/07/2021
Initial Date FDA Received01/20/2021
Supplement Dates Manufacturer Received02/16/2021
03/10/2021
05/18/2021
Supplement Dates FDA Received03/09/2021
03/10/2021
05/18/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
ALIN I SYPHILS 1200T CN, 07P60-77, (B)(6); ALNTY I PROCESSING MODU, 03R65-01,; ALNTY I PROCESSING MODU, 03R65-01, (B)(6).; ALNTY I PROCESSING MODU, 03R65-01, (B)(6).
Patient Age59 YR
-
-