Model Number 16-02-80
Device Problem Microbial Contamination of Device (2303)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
There was no known patient involvement.The heater-cooler 16-02-80 is not distributed in the usa and it is similar to heater-cooler 16-02-85, which is distributed in the usa (510(k) number: k191402).Livanova (b)(4) implemented a field safety notice for disinfection and cleaning of heater-cooler devices.The z number is z-2076/2081-2015.Livanova (b)(4) manufactures the heater-cooler system 3t.The incident occurred in (b)(6).A review of the dhr did not identify any deviations or non-conformities relevant to the reported issue.Through follow up communication under a previous case from the same hospital, livanova (b)(4) learned that the device was cleaned regularly per the ifu, placed inside of the operation theater during use, with the fan facing away from the patient in direction of an exhaust fan in the wall.The device was kept in a housing with negative pressure in relation to the pressure in the or.The estimated distance between the surgery field and the device is 3 meters.If any additional information is received, a follow up report will be submitted.
 
Event Description
Livanova (b)(4) received report (b)(4) from (b)(6) stating that a heater-cooler system 3t device was found to be contaminated with mycobacterium paragordonae (sample from (b)(6) 2020).The laboratory report confirming the reported contamination has been provided to livanova.There is no known patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEATER-COOLER SYSTEM 3T
Type of Device
CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich 80309
GM  80309
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich 80309
GM   80309
Manufacturer Contact
enrico greco
14401 w. 65th way
arvada 80004
MDR Report Key11207076
MDR Text Key227999486
Report Number9611109-2021-00036
Device Sequence Number1
Product Code DWC
Combination Product (y/n)N
Reporter Country CodeSZ
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Other Health Care Professional
Remedial Action Recall
Type of Report Initial
Report Date 01/18/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number16-02-80
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/22/2020
Initial Date FDA Received01/21/2021
Date Device Manufactured03/10/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberZ-2076/2081-2015
Patient Sequence Number1
-
-