Model Number 720054-02 |
Device Problem
Inadequacy of Device Shape and/or Size (1583)
|
Patient Problems
No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/29/2020 |
Event Type
Injury
|
Event Description
|
It was reported that the patient went to the hospital two weeks before surgery, saying the spectra penile prosthesis length was not correct.As a result of the inspection, it was confirmed that the length was not correct.The patient had the spectra penile prosthesis replaced and extended the tip by 1.5 cm.Following the surgery, the patient was stable and improved.The event resolved.
|
|
Manufacturer Narrative
|
H3 device evaluation the rear tip extender was not returned for analysis.The spectra cylinders were visually and functionally tested.Both cylinders had sharp instrument/tool damage to the outer tube, which resulted in a hole and exposed metal segments in cylinder body.This damage is consistent with explant damage.Both cylinders passed the bend test with a force readout of less than 1390 grams.The cylinders therefore performed within specification.Product analysis was unable to confirm the reported event.
|
|
Event Description
|
It was reported that the patient went to the hospital two weeks before surgery, saying the spectra penile prosthesis length was not correct.As a result of the inspection, it was confirmed that the length was not correct.The patient had the spectra penile prosthesis replaced and extended the tip by 1.5 cm.Following the surgery, the patient was stable and improved.The event resolved.
|
|
Search Alerts/Recalls
|