• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS MEDICAL SYSTEMS CORP. EVIS EXERA II VIDEO SYSTEM CENTER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS MEDICAL SYSTEMS CORP. EVIS EXERA II VIDEO SYSTEM CENTER Back to Search Results
Model Number CV-180
Device Problem Image Display Error/Artifact (1304)
Patient Problems No Patient Involvement (2645); No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
The device was returned, and an initial evaluation was conducted by olympus; however, investigation is ongoing.If additional information becomes available following device evaluation, a supplemental report will be filed.
 
Event Description
The evis exera ii video system was returned to olympus for routine inspection.During this inspection it was discovered that the front panel was faded and a loose latch was causing a flickering display.There was no patient involvement for this event.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer¿s final investigation.As a result of the physical inspection and functional testing of the device, olympus has concluded that the locking lever failed due to deterioration from repeated use over a long period of time.It is speculated that this resulted in an unstable connection of the video connectors and instability of the electrical contacts of the connectors, resulting in flickering of the images.Based on the legal manufacturer¿s investigation, the device history record (dhr) was reviewed and there were no problems found during the manufacturing of the device that would cause or contribute to the reported issue.The device met all specifications at the time of shipment.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS EXERA II VIDEO SYSTEM CENTER
Type of Device
VIDEO SYSTEM CENTER
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
MDR Report Key11215620
MDR Text Key228240845
Report Number8010047-2021-01739
Device Sequence Number1
Product Code FAJ
UDI-Device Identifier04953170215513
UDI-Public04953170215513
Combination Product (y/n)N
PMA/PMN Number
K133538
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup
Report Date 03/11/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCV-180
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/30/2020
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/30/2020
Initial Date FDA Received01/22/2021
Supplement Dates Manufacturer Received03/05/2021
Supplement Dates FDA Received03/11/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-