Model Number CV-180 |
Device Problem
Image Display Error/Artifact (1304)
|
Patient Problems
No Patient Involvement (2645); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Manufacturer Narrative
|
The device was returned, and an initial evaluation was conducted by olympus; however, investigation is ongoing.If additional information becomes available following device evaluation, a supplemental report will be filed.
|
|
Event Description
|
The evis exera ii video system was returned to olympus for routine inspection.During this inspection it was discovered that the front panel was faded and a loose latch was causing a flickering display.There was no patient involvement for this event.
|
|
Manufacturer Narrative
|
This report is being supplemented to provide additional information based on the legal manufacturer¿s final investigation.As a result of the physical inspection and functional testing of the device, olympus has concluded that the locking lever failed due to deterioration from repeated use over a long period of time.It is speculated that this resulted in an unstable connection of the video connectors and instability of the electrical contacts of the connectors, resulting in flickering of the images.Based on the legal manufacturer¿s investigation, the device history record (dhr) was reviewed and there were no problems found during the manufacturing of the device that would cause or contribute to the reported issue.The device met all specifications at the time of shipment.
|
|