SMITH & NEPHEW, INC. LIGHT SOURCE 500XL XENON; IMAGE, ILLUMINATION, FIBEROPTIC, FOR ENDOSCOPE
|
Back to Search Results |
|
Model Number 72200568 |
Device Problems
Image Display Error/Artifact (1304); Overheating of Device (1437)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/06/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
Internal complaint reference: (b)(4).
|
|
Event Description
|
It was reported that, during a procedure, the light source 500xl xenon displayed an e13 error on the screen.The procedure was completed with the same device.It is unknown if there was a delay.No patient injury or other complications were reported.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
|
|
Manufacturer Narrative
|
The reported device, used in treatment, was not returned to the designated complaint unit for independent evaluation, thus visual inspection and functional testing could not be performed.There was a relationship found between the returned device and the reported incident.A review of the customer-supplied photograph was performed and observed an e13 error code.The complaint was confirmed and the root cause was associated with component failure.Factors, unrelated to the manufacturing and design of the device that could have contributed to the reported event, include overheating of the power supply caused by poor ventilation due to external obstructions to vents or excessive dust buildup on rear fans or fan guards.A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A complaint history review concluded this was a repeat issue.
|
|
Search Alerts/Recalls
|
|
|