Based on the images provided of the incident, the complaint can be confirmed.Argon has recalled tlab products for the 7f introducer sheath distal tip separating during procedures.Lot number 1451428 used in this incident was included in the recall that went into effect december 2020.It is unknown at this time why the complainant facility still had the product in its possession and used after the recall notifications were sent to customers.The manufacturing defect was identified before argon started receiving the complaints that prompted the recall.(b)(4) was performed and it adequately addressed the root causes and corrective/preventative actions taken.All affected lots were manufactured before the tip bonding process was updated.Corrective actions included modifying the tipping mandrels to shorten the narrowed tip to 0.300¿, adding band markings at 0.100¿ and 0.200¿ onto the mandrel to simplify the locating of the mandrel within the introducer sheath tube and aid in the alignment of the tube and mandrel within the tipping die by the operator.It also included adding a 100% bend test as well as 100% visual inspection for wrinkling to the inspection process.
|