RESPIRONICS CALIFORNIA, INC V60 VENTILATOR; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE
|
Back to Search Results |
|
Model Number V60 |
Device Problem
Key or Button Unresponsive/not Working (4063)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Date of event: (b)(6) 2020.Date of report: 25jan2021.
|
|
Event Description
|
It was reported to philips that the device had a navigation ring that was not responding.The device was not in clinical use at the time of the event.This issue was found during set up before use.There was no delay to patient care and a different device was used.There was no report of patient or user harm.The customer requested that a philips field service engineer (fse) be dispatched to the customer site to provide additional assistance.The fse replaced the front bezel assembly to resolve the reported issue.The device successfully passed testing and was returned to service.
|
|