BOSTON SCIENTIFIC CORPORATION ACCOLADE MRI EL DR; IMPLANTABLE PULSE GENERATOR, PACEMAKER (NON-CRT)
|
Back to Search Results |
|
Model Number L331 |
Device Problems
Signal Artifact/Noise (1036); High impedance (1291); Failure to Sense (1559); Connection Problem (2900)
|
Patient Problems
No Known Impact Or Consequence To Patient (2692); Insufficient Information (4580)
|
Event Date 01/01/2013 |
Event Type
Injury
|
Event Description
|
It was reported that this pacemaker system triggered a lead safety switch (lss) due to a high out of range right ventricular (rv) lead pace impedance measurement of greater than 2000 ohms.Additionally, there was noise present on the rv lead, but it was not sensed by the device therefore, no pacing inhibiting occurred.Technical services (ts) was consulted and offered troubleshooting and programming options.Upon further review, the rv lead impedance measurement have been gradually increasing over the past two years.Patient was brought into the clinic for further follow up and device reprogramming.At this time this pacemaker system remains in service.No adverse patient effects were reported.
|
|
Event Description
|
It was reported that this pacemaker system triggered a lead safety switch (lss) due to a high out of range right ventricular (rv) lead pace impedance measurement of greater than 2000 ohms.Additionally, there was noise present on the rv lead, but it was not sensed by the device therefore, no pacing inhibiting occurred.Technical services (ts) was consulted and offered troubleshooting and programming options.Upon further review, the rv lead impedance measurement have been gradually increasing over the past two years.Patient was brought into the clinic for further follow up and device reprogramming.Subsequently, the patient underwent an rv lead revision procedure and during that procedure, while attempting to re-insert the right atrial (ra) lead into the device header, it was noted upon connection, the atrial electrogram (egm) did not display, therefore unable to sense.The device was successfully explanted and replaced.The rv lead was surgically abandoned and replaced.The ra lead remains in service with appropriate sensing reported at this time.No adverse patient effects were reported.
|
|
Manufacturer Narrative
|
The returned device was thoroughly inspected and analyzed upon receipt at our post market quality assurance laboratory.Visual examination of the device header and case noted no anomalies.Pin gauge testing, designed to verify proper port dimensions, was completed.Each port measured as expected.The device was then exposed to simulated heart load conditions, and the pacing, and sensing functions were tested.Impedance testing was completed and all measurements were within normal limits.The device operated appropriately with no interruptions in therapy output at the returned programmed settings.A series of electrical tests was also performed, and again, normal device function was observed.Analysis did not identify any device characteristics that would have caused or contributed to the reported clinical observations.
|
|
Event Description
|
It was reported that this pacemaker system triggered a lead safety switch (lss) due to a high out of range right ventricular (rv) lead pace impedance measurement of greater than 2000 ohms.Additionally, there was noise present on the rv lead, but it was not sensed by the device therefore, no pacing inhibiting occurred.Technical services (ts) was consulted and offered troubleshooting and programming options.Upon further review, the rv lead impedance measurement have been gradually increasing over the past two years.Patient was brought into the clinic for further follow up and device reprogramming.Subsequently, the patient underwent an rv lead revision procedure and during that procedure, while attempting to re-insert the right atrial (ra) lead into the device header, it was noted upon connection, the atrial electrogram (egm) did not display, therefore unable to sense.The device was successfully explanted and replaced.The rv lead was surgically abandoned and replaced.The ra lead remains in service with appropriate sensing reported at this time.No adverse patient effects were reported.
|
|
Search Alerts/Recalls
|
|
|