• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LIVANOVA DEUTSCHLAND HEATER-COOLER SYSTEM 3T; CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LIVANOVA DEUTSCHLAND HEATER-COOLER SYSTEM 3T; CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 16-02-85
Device Problem Electrical /Electronic Property Problem (1198)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/28/2020
Event Type  malfunction  
Manufacturer Narrative
There was no patient involvement.Livanova (b)(6) manufactures the heater-cooler system 3t.The incident occurred in (b)(6).A livanova field service representative was dispatched to the facility to investigate the device and no error could be reproduced and the device was found to be properly working.However, the technician decided to replace the ac mains board.Subsequent functional verification testing was completed without further issues and the unit was returned to service.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Event Description
Livanova (b)(6) received a report that a heater-cooler system 3t displayed an error code on both patient and cardioplegia circuits during procedure.The user did not remember the error code.The device was replaced with a back-up to complete the procedure.There was no report of patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEATER-COOLER SYSTEM 3T
Type of Device
CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich 80309
GM  80309
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich 80309
GM   80309
Manufacturer Contact
enrico greco
14401 w. 65th way
arvada 80004
MDR Report Key11228772
MDR Text Key231146993
Report Number9611109-2021-00042
Device Sequence Number1
Product Code DWC
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K191402
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Other Health Care Professional
Remedial Action Other
Type of Report Initial
Report Date 01/25/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number16-02-85
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/28/2020
Initial Date FDA Received01/26/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/10/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-