Device evaluated by mfr.: the product was returned to boston scientific for analysis.Returned product consisted of an fractionated flow reserve comet pressure wire connected to the optical cable.The tip, device shaft, sensor port and the coefficient values were examined for damage or any irregularities.The shaft showed 3 kinks located 33cm, 38cm and 136.5cm from the tip.The tip showed a bend and stretched coils.There was scraped coating at the 33cm location.The optical cable handle was connected to the fractionated flow reserve link to verify the signal strength.The signal was not present as designed.The coefficient was confirmed to be programmed.The optical cable handle cap was loosened to remove the wire.There was no issue with removing the wire.The sensor housing showed no residue of body fluids.Inspection of the remainder of the device, apart from the observed damage, revealed no other damage or irregularities.Device analysis determined the condition of the returned device was consistent with the reported information.
|