MAUDE Adverse Event Report: ABBOTT VASCULAR MITRACLIP XTR CLIP DELIVERY SYSTEM; VALVE REPAIR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR MITRACLIP XTR CLIP DELIVERY SYSTEM; VALVE REPAIR Back to Search Results
Model Number CDS0601-XTR
Device Problem Incomplete Coaptation (2507)
Patient Problems Dyspnea (1816); Edema (1820); Mitral Regurgitation (1964)
Event Date 01/18/2021
Event Type  Injury  
Manufacturer Narrative
The device was not returned for analysis.A review of the lot history record revealed no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no similar incidents reported from this lot.Based on the available information, a cause for the reported single leaflet device attachment (slda) and poor image resolution could not be determined.The reported patient effects of dyspnea, edema, and recurrent mitral regurgitation (mr) appear to have been cascading events of the reported slda.The reported patient effects of dyspnea, edema, and recurrent mr as listed in the mitraclip system instructions for use, are known possible complications associated with mitraclip procedures.There is no indication of a product quality issue with respect to manufacture, design or labeling.The additional device referenced is filed under a separate medwatch report number.
 
Event Description
This is being filed to report the clip detaching from one leaflet and the mr increased.It was reported that the initial mitraclip procedure was performed on (b)(6) 2019, to treat degenerative mitral regurgitation (mr) with a grade of 4.Two clips were implanted, reducing mr to 2.It was noted echo quality was suboptimal.On (b)(6) 2021, the patient presented with shortness of breath and edema of the legs.It was noted both clips had detached from the posterior leaflet (single leaflet device attachment (slda)) and the mr increased to 3-4.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MITRACLIP XTR CLIP DELIVERY SYSTEM
Type of Device
VALVE REPAIR
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 3005070406
3885 bohannon drive
menlo park CA 94025
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key11230153
MDR Text Key228675517
Report Number2024168-2021-00616
Device Sequence Number1
Product Code NKM
UDI-Device Identifier08717648226366
UDI-Public08717648226366
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P100009
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 01/26/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/13/2020
Device Model NumberCDS0601-XTR
Device Catalogue NumberCDS0601-XTR
Device Lot Number90814U190
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/18/2021
Initial Date FDA Received01/26/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/14/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
1 IMPLANTED MITRACLIP
Patient Outcome(s) Other;
Patient Age68 YR
Patient Weight78
-
-