Additional information was added to h3, h4, and h6.H4: the lot was manufactured from august 11, 2020 - august 12, 2020.H10: four (4) samples were received for evaluation.A visual inspection was performed, and it was noted that the spikes of 2 samples were detached and found loose in the opened primary packaging.Additionally, the detached spikes of the other 2 samples were missing.The reported condition was verified.The cause of the condition could not be determined; however, the most likely cause was due to inadequate or lack of adhesive being applied in the manufacturing process at the spike connection to the tubing.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|