The sensor was investigated, and customer's complaint was reproduced via in-vitro qc test and was also confirmed during review of the in-vivo data.The issue is potentially due to an unacceptable high sensitivity of the off channels of this sensor to changes in temperature.The investigation shows the system correctly disabled the sensor due to performance failure and the system's self-test functions are working normally.D9 device available for evaluation? yes, device received on 02 feb 2021.H3.Device evaluated by manufacturer? yes.H6.Type of investigation updated to 10; h6.Investigation findings updated to 3231; h6.Investigation conclusions updated to 4307.
|