• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SENSEONICS INC. EVERSENSE SENSOR; EVERSENSE CONTINUOUS GLUCOSE MONITOR SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SENSEONICS INC. EVERSENSE SENSOR; EVERSENSE CONTINUOUS GLUCOSE MONITOR SYSTEM Back to Search Results
Model Number 102096-67A
Device Problem Device Sensing Problem (2917)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/29/2020
Event Type  malfunction  
Manufacturer Narrative
The manufacturer is currently performing an investigation and will provide the results with the supplemental report.
 
Event Description
On (b)(6) 2020,senseonics was made aware of an incident where customer complaints of getting high ambient alerts resulting in an early sensor removal.
 
Manufacturer Narrative
The sensor was investigated, and customer's complaint was reproduced via in-vitro qc test and was also confirmed during review of the in-vivo data.The issue is potentially due to an unacceptable high sensitivity of the off channels of this sensor to changes in temperature.The investigation shows the system correctly disabled the sensor due to performance failure and the system's self-test functions are working normally.D9 device available for evaluation? yes, device received on 02 feb 2021.H3.Device evaluated by manufacturer? yes.H6.Type of investigation updated to 10; h6.Investigation findings updated to 3231; h6.Investigation conclusions updated to 4307.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVERSENSE SENSOR
Type of Device
EVERSENSE CONTINUOUS GLUCOSE MONITOR SYSTEM
Manufacturer (Section D)
SENSEONICS INC.
20451 seneca meadows parkway
germantown MD 20875 7005
MDR Report Key11238479
MDR Text Key229108057
Report Number3009862700-2021-00012
Device Sequence Number1
Product Code QCD
UDI-Device Identifier00817491022349
UDI-Public00817491022349
Combination Product (y/n)Y
PMA/PMN Number
P160048
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Type of Report Initial,Followup
Report Date 12/29/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date12/11/2020
Device Model Number102096-67A
Device Catalogue NumberFG-4200-00-301
Device Lot NumberWP07612
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/02/2021
Initial Date Manufacturer Received 12/29/2020
Initial Date FDA Received01/27/2021
Supplement Dates Manufacturer Received12/29/2020
Supplement Dates FDA Received02/24/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age26 YR
-
-