Model Number 60-02-15 |
Device Problem
Pumping Problem (3016)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 12/25/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
There was no patient involvement.Livanova (b)(4) manufactures the centrifugal pump system with tubing clamp.The incident occurred in (b)(6).The reported issue could be confirmed through provided videos.A livanova field service representative was dispatched to the facility to investigate and found that the mounting nut of the knob was loose causing the reported failure.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
|
|
Event Description
|
Livanova (b)(4) received a report of a centrifugal pump system with tubing clamp with the speed control knob which could not be rotated during use.There was no report of patient injury.
|
|
Manufacturer Narrative
|
H.10: a review of the dhr could not identify any deviations or non conformities relevant to the issue.The knob nut was found to be loose and the technician re-tightened it in order to solve the problem.Subsequently functional tests were preformed and passed and the unit was returned to regular service.
|
|
Event Description
|
See initial report.
|
|