Model Number 60-02-15
Device Problem Pumping Problem (3016)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/25/2020
Event Type  malfunction  
Manufacturer Narrative
There was no patient involvement.Livanova (b)(4) manufactures the centrifugal pump system with tubing clamp.The incident occurred in (b)(6).The reported issue could be confirmed through provided videos.A livanova field service representative was dispatched to the facility to investigate and found that the mounting nut of the knob was loose causing the reported failure.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Event Description
Livanova (b)(4) received a report of a centrifugal pump system with tubing clamp with the speed control knob which could not be rotated during use.There was no report of patient injury.
 
Manufacturer Narrative
H.10: a review of the dhr could not identify any deviations or non conformities relevant to the issue.The knob nut was found to be loose and the technician re-tightened it in order to solve the problem.Subsequently functional tests were preformed and passed and the unit was returned to regular service.
 
Event Description
See initial report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CENTRIFUGAL PUMP SYSTEM WITH TUBING CLAMP
Type of Device
CONTROL, PUMP SPEED, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich
MDR Report Key11242915
MDR Text Key240342299
Report Number9611109-2021-00048
Device Sequence Number1
Product Code DWA
Combination Product (y/n)N
PMA/PMN Number
K032213
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Remedial Action Other
Type of Report Initial,Followup
Report Date 01/25/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number60-02-15
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/29/2020
Initial Date FDA Received01/28/2021
Supplement Dates Manufacturer Received02/02/2021
Supplement Dates FDA Received03/03/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-