This follow-up report is being submitted to relay additional information.The following sections were updated: b4, b5, g4, g7, h1, h2, h10 visual examination of the returned product no visible damage on the returned device.Functional testing of the sleeve found the sleeve functioned as intended.Review of the product determined that no failure was found as the product functions as intended.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|