• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. JGRKNT 1.0MM MINI 2-0 NDLS; FASTNER, FIXATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. JGRKNT 1.0MM MINI 2-0 NDLS; FASTNER, FIXATION Back to Search Results
Catalog Number 912076
Device Problems Material Twisted/Bent (2981); Mechanical Jam (2983)
Patient Problem No Information (3190)
Event Date 10/26/2020
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Report source: (b)(6).Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up report will be submitted.
 
Event Description
It was reported that the surgeon was unable to move the sleeve on the juggerknot.
 
Event Description
No further event information available at the time of this report.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.A juggerknot was returned for evaluation.A functional check of the sleeve was performed and was conforming to specifications.Visual examination of the returned product identified the tip to bent and it has likely happened during the surgery.Device history record was reviewed and no discrepancies were found.No failure was detected as the sleeve was working as intended.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JGRKNT 1.0MM MINI 2-0 NDLS
Type of Device
FASTNER, FIXATION
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
MDR Report Key11247343
MDR Text Key230730664
Report Number0001825034-2021-00282
Device Sequence Number1
Product Code DZL
Combination Product (y/n)N
PMA/PMN Number
K110879
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 03/03/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number912076
Device Lot Number352930
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/27/2021
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 01/04/2021
Initial Date FDA Received01/28/2021
Supplement Dates Manufacturer Received03/02/2021
Supplement Dates FDA Received03/03/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-