Model Number 12TLW404F
Device Problem Material Rupture (1546)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/07/2020
Event Type  malfunction  
Manufacturer Narrative
One 12tlw404f catheter was returned for examination.The reported event of ¿ruptured¿ was confirmed.The balloon was found to be ruptured between the windings.The distal balloon winding was removed to match the latex edges and the latex edges did not appear to match at the ruptured region.No other visible damage was observed from the catheter body and windings.The through lumen was patent without any leakage or occlusion.A review of the manufacturing records indicated that the product met specifications upon release.An investigation has been initiated to consider any potential manufacturing factors that may have contributed to this complaint.Balloon rupture and catheter separation as a result of excessive pull force applied to remove adherent material are the most frequent causes of reported failures.To minimize the risk of vessel damage, balloon rupture, or tip detachment, the maximum recommended inflation volume and pull force for the catheter should not be exceeded.The thru-lumen embolectomy catheter is not recommended for the removal of fibrous, adherent, or calcified material (e.G.Chronic clot, atherosclerotic plaque).This catheter is not designed to withstand the additional pull force needed to remove these materials.The instructions for use of the product contains the following statement: ¿as with all catheterization procedures, complications may occur.These may include local or systemic infection, local hematomas, intimal disruption, arterial dissection, perforation and vessel rupture, hemorrhage, arterial thrombosis, distal embolization of blood clots and atherosclerotic plaque, air embolus, aneurysm, arterial spasm, arteriovenous fistula formation, and balloon rupture with fragmentation, tip separation and distal embolization.¿ complaint histories for all reported events are reviewed against trending control limits on a monthly basis, and any excursions above the control limits are assessed and documented as part of this monthly review.Udi: (b)(4).
 
Event Description
It was reported that the balloon of a fogarty catheter ruptured during a thrombectomy procedure.No patient complications were reported.Patient demographics were unable to be obtained.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FOGARTY THRU-LUMEN EMBOLECTOMY CATHETER
Type of Device
CATHETER, EMBOLECTOMY
Manufacturer (Section D)
EDWARDS LIFESCIENCES PR
state rd indus pk 402 km 1.4
anasco PR
Manufacturer (Section G)
EDWARDS LIFESCIENCES PR
state rd indus pk 402 km 1.4
anasco PR
Manufacturer Contact
jessica atallah
1 edwards way
irvine, CA 92614
9492500294
MDR Report Key11248005
MDR Text Key232590529
Report Number2015691-2021-00243
Device Sequence Number1
Product Code DXE
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K892410
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 12/07/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/25/2022
Device Model Number12TLW404F
Device Catalogue Number12TLW404F
Device Lot Number63349788
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/07/2021
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/12/2021
Initial Date FDA Received01/28/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/31/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-