Updated fields: b4, g3, g6, h2, h6, h10.The suspected faulty pressure transducer was inspected by a sr.Electronic technician at the sustaining engineering, and no visual defects were found during visual inspection.In order to reproduce the failure, the suspect pressure transducer was installed in the cs100 test fixture, and the following tests were perform per procedure: volume test, safety disk leak test, k6, k6a, k7, k8 leak test, safety vent test, pneumatic performance test, autofill calibration.All tests passed.The sr.Electronic technician then ran the test fixture that was pumped at 120bpm for 24 hours and no failures were reported.The sr.Electronic technician at the sustaining engineering was unable to duplicate the reported failure.Per procedure, the pressure transducer will be discarded.
|