Device inspection found a broken suction port.Device is non repairable.The investigation is ongoing; therefore, the root cause of the reported event cannot be determined at this time.However, if additional information becomes available this report will be supplemented accordingly.
|
This report is being supplemented to provide additional information based on the legal manufacturer¿s final investigation.The device has been previously serviced by olympus, therefore a service history review will replace dhr review.This device has undergone multiple inspections as being asset returns.This device was previously inspected by olympus in january 2020 (ds# 20006886).At that time, minor scratches on the housing were observed.The unit passed functional test, output test, and electrical safety test.Based on the device inspection result, damage to the device was most likely due to user's mishandling.The ifu advises "this product is a precision device; handle it with care.Avoid rough or violent handling, which may cause equipment damage.".
|