RESPIRONICS CALIFORNIA, INC V60 VENTILATOR; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE
|
Back to Search Results |
|
Model Number V60 |
Device Problems
Unexpected Shutdown (4019); Application Program Freezes, Becomes Nonfunctional (4031)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Date of event: (b)(6) 2021.Date of report: 29jan2021.
|
|
Event Description
|
The customer reported a ventilator freezes or shuts down when switching to high flow therapy (hft) during set up.The device was evaluated by the customer and a remote service engineer (rse).The device was not in clinical use at the time the issue was discovered.There was no patient or user harm reported.There was no delay to patient therapy reported.The rse walked the customer through changing the ventilator mode from standby to high flow therapy (hft) with no issues.Additionally, upon reviewing the event log a circuit occlusion error code was found.It was determined the error code was caused by a circuit occlusion.
|
|